Enviro Test Construct LCD Laboratory Conductivity Meter

This LCD Laboratory Conductivity Meter has 0.01, 0.1, 1 salinity resolution, 0.01, 0.1, 1 TDS resolution, and full range of ±1% salinity accuracy.

SKU: ETCLCM-217 Category:

Technical Specifications

Electric conductivity Measuring Range0.01 µS/cm to 20.00 µS/cm, 200.0 µS/cm, 2000 µS/cm, 20.00 mS/cm, 200.0 mS/cm
Display Resolution0.001, 0.01, 0.1, 1
Measurement AccuracyFull range ±0.5%
Calibration PointOne to three o ‘clock
Use of Calibration Fluid TDS10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm
Measuring Range0 mg/L to 10.00 mg/L, 100.0 mg/L, 1000 mg/L, 10.00 g/L, 200.0 g/L
Display Resolution0.01, 0.1, 1
Measurement AccuracyFull range ±1%
TDS Conversion Factor0.1 to 1.0 (default: 0.5)
Salinity Measuring Range0.00 ppt to 80.00 ppt, 0.00 psu to 42.00 psu, 0.00% to 8.00%
Display Resolution0.01, 0.1, 1
Measurement AccuracyFull range ±1%
Electrical Resistivity Measuring Range 0.00 MΩ to 30.00 MΩ
Display Resolution0.01, 0.1, 1
Measurement AccuracyFull range ±1%
Conductivity Measuring Range0% to 00%
Display Resolution0.01, 0.1, 1
Measurement AccuracyFull range ±1%
Measurement ModeRefined sugar or raw sugar
Temperature Compensation CoefficientLinear (0.0% to 10.0% / °C), non-linear, pure water
Conductivity Cell Constant2-pole (K=0.1, 1, 10) or 4-ring conductance electrode
Stability ConditionStandard or high precision
Calibration Expiration Reminder1 day to 31 days or closed
Time Interval Reading10 seconds, 30 seconds, 60 seconds, 10 minutes, 30 minutes or off
Password ProtectionFour-digit number
Data Storage1000 groups
Communication OutputUSB
Coupler6-pin DIN, 0.13 in (3.5 mm) microphone socket
Display Screen7-in TFT LCD display
Power Supply TypeDC 12 V power adapter
Reference Temperature68 °F or 77 °F (20 °C or 25 °C)
Temperature Compensation Range32 °F to 212 °F (0 ℃ to 100 ℃), manual or automatic
Temperature Measuring Range32 °F to 212 °F (0 ℃ to 100 ℃)
Display Resolution32.18 °F (0.1 ℃)
Measurement Accuracy±0.9 °F (±0.5 °C)
Offset Calibration1 point, measuring ±50 °F (±10 °C)
Overall Dimension9.45 in x 8.66 in x 3.15 in (240 mm x 220 mm x 80 mm)
Weight3.75 lb. (1.7 kg)

Reviews

There are no reviews yet.

Be the first to review “Enviro Test Construct LCD Laboratory Conductivity Meter”

Your email address will not be published. Required fields are marked *