Conductivity Controller with TDS Measuring (Temp. Alarm)

SKU: ETC00YZ00ZB Category:

Key Features

  • Economical online conductivity controller contains TDS measurement mode.
  •  1 to 3 points calibration with automatic calibration solution recognition.
  •  Selectable cell constant, temperature coefficient and TDS conversion factor.
  •  Automatic Temperature Compensation ensures accurate reading.
  •  Low and high alarm limits can be set by user and automatically activate an external device.
  • 4 to 20mA output signal meets industry standard.
  •  Manual temperature calibration improves the measurement accuracy of meter.
  •  Selectable temperature unit (°C or °F) for different application requirements.
  • System menu allows setting the 9 parameters, including the number of calibration points,
    cell constant, alarm limits, hysteresis value, etc.
  •  Reset feature automatically resumes all settings back to factory default settings.

Technical Specifications

Conductivity
Range0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
Accuracy±1% F.S
Resolution0.01, 0.1, 1
Calibration Points1 to 3 points
Calibration Solutions84uS/cm, 1413uS/cm, 12.88mS/cm, 111.8mS/cm
TDS
Range0~100ppt (Max. 200ppt, depending on factor setting)
Accuracy±1% F.S
TDS Factor0.1~1.0 (Default 0.5)
Temperature
Range0~100°C, 32~212°F
Accuracy±1°C
Resolution0.1°C
Communication
Signal Output4~20mA
LoadMax. 500Ω
Low and high alarm limits0.02uS/cm~20.0mS/cm, Selectable
Communication InterfaceRS485
Connection TerminalsDetachable screw terminals
Others
Temperature Compensation0~100°C, 32~212°F, Automatic
Temperature Coefficient0.0~10.0%/°C
Cell ConstantK=0.1, 1, 10
Normalization Temperature25°C
Environmental Conditions0~60°C, 32~140°F
Power RequirementsDC24V
Dimensions3.78in x 3.78in x 2.92in

(96(L)×96(W)×75(H)mm)

Weight 0.77lbs  (350g)

Reviews