Enviro Test Construct Advanced Multifunction Temperature and Signal Calibrator is a portable field calibrator powered by a large-capacity high-performance polymer rechargeable battery. It can be connected to an external pressure module. It is easy to operate, powerful, stable, and highly accurate. It is very cost-effective. High thermal comprehensive calibration instrument.

 

Features

  • It can output V, mV, mA, Ω, and also can simulate thermocouple or thermal resistance signals;
  • It can measure V, mV, mA, Ω, pressure, and calibrating pressure switch, and also can use a thermocouple or thermal resistance signals to measure temperature;
  • Measurement and output can be carried out at the same time, and mutually separated;
  • It provides 24V DC power, and separates from measurement and output;
  • High-definition display screen, favorable human-machine interface, full-Chinese operation interface;
  • It supports step output, and can automatically program commonly-used step signals and step intervals, and it supports two modes of automatic steps and manual steps;
  • It adopts a Pt100 temperature measurement probe for cold-end temperature compensation, and supports two modes of ambient temperature tracking and constant temperature setting, which will facilitate use;
  • With multi-function and programmable unit conversion function, it can converse V, mA output or measurement signals into other units such as °C, °F, %, kPa, MPa, psi, mmHg, mmH₂O, m³/s, m³/min, m³/h, L/s, L/h, L/min, m, ?;
  • When outputting mA signals, it supports two modes of mA source and simulation transmitter;
  • When outputting and measuring thermocouple °C signal, it displays corresponding mV value;
  • When outputting and measuring thermal resistance °C signal, it displays corresponding Ω value;
  • When outputting and measuring Ω signal, it supports two-, three-, and four-line modes;
  • It provides 24V DC loop source, and measures mA signals in the loop;
  • It can conduct signal fine-tuning at any digit of output value;
  • When measuring values, it can lock or unlock measurement values;
  • When measuring values, it can collect minimum and maximum values, and calculate mean value.

 

Technical Specifications

Display3.7-inch high-definition black-white screen
Output and Measurementφ4.0mm gold-plated jack
24V DC7-core aviation plug
Pressure Communication7-core aviation plug
Cold-End Temperature Compensation6-core aviation plug
Switch Detection6-core aviation plug
PC Communication5-core aviation plug/RS232-DB9 socket
RS232 ParametersBaud rate 9600; data length of 8 bits; stop bit of 1 bit
HART Communication5-core aviation plug
Charging Jackφ2.1mm charging socket
Real-Time ClockLayout of [YYMMDD], year setting range is 2000 ~ 2099, and 24-hour time system is adopted.
Dimension (LxWxH)4.3 in x 8.07 in x 2.04 in (11 cm  × 20.5 cm  × 5.2 cm)
Weight31.74 oz (About 900 g)

 

  • Output Technical Indicators (One Year, Ambient Temperature: 20 ± 5°C)
Output Range0.01-Grade Max. Error Allowed0.02-Grade Max. Error Allowed0.05-Grade Max. Error AllowedDefinition
10.99999V± (0.008% reading + 0.0001V)± (0.015% reading + 0.0001V)± (0.035% reading + 0.0001V)10μV
★ 1.099999V± (0.008% reading + 0.0001V)± (0.015% reading + 0.0001V)± (0.035% reading + 0.0001V)1μV
-99.9999mV to 109.9999mV± (0.008% reading + 0.003mV)± (0.015% reading + 0.003mV)± (0.035% reading + 0.003mV)0.1μV
10 Types of Thermocouple °CSee “TC Thermocouple Technical Indicators”0.1°C
★ 30.0999mA± (0.01% reading + 1μA)± (0.015% reading + 1μA)± (0.035% reading + 1μA)0.1μA
4000.00Ω± (0.008% reading + 0.04Ω)± (0.015% reading + 0.04Ω)± (0.035% reading + 0.04Ω)10mΩ
★ 400.000Ω± (0.008% reading + 0.005Ω)± (0.015% reading + 0.005Ω)± (0.035% reading + 0.005Ω)1mΩ
6 Types of Thermal Resistance °CSee “RTD Thermal Resistance Technical Indicators”0.01°C
24V DC Loop Power± 5%± 5%± 5%
V max. output current: 10mA; 20mA output loading capacity ≤ 1kΩ; 5000Hz loading capacity ≥ 10kΩ; 24V DC loop power max. output current: 80mA Resistance, thermal resistance external current working range: 400Ω: (0.5 ~ 7.5)mA; 4000Ω: (0.05 ~ 0.75)mA

 

  • Measurement Technical Indicators (One Year, Ambient Temperature: 20 ± 5°C)

 

Measurement Range0.01-Grade Max. Error Allowed0.02-Grade Max. Error Allowed0.05-Grade Max. Error AllowedDefinition
± 59.9999V± (0.008% reading + 0.0005V)± (0.015% reading + 0.0005V)± (0.035% reading + 0.0005V)0.1mV
★ ± 5.99999V± (0.008% reading + 0.00005V)± (0.015% reading + 0.00005V)± (0.035% reading + 0.00005V)0.01mV
± 599.999mV± (0.008% reading + 0.005mV)± (0.015% reading + 0.005mV)± (0.035% reading + 0.005mV)1μV
± 119.999mV± (0.008% reading + 0.003mV)± (0.015% reading + 0.003mV)± (0.035% reading + 0.003mV)1μV
10 Types of Thermocouple °CSee “TC Thermocouple Technical Indicators”0.1°C
★ ± 119.999mA± (0.01% reading + 1μA)± (0.015% reading + 1μA)± (0.035% reading + 1μA)1μA
★ ± 23.9999mA± (0.01% reading + 1μA)± (0.015% reading + 1μA)± (0.035% reading + 1μA)0.1μA
5999.99Ω± (0.008% reading + 0.05Ω)± (0.015% reading + 0.05Ω)± (0.035% reading + 0.05Ω)10mΩ
★ 599.999Ω± (0.008% reading + 0.005Ω)± (0.015% reading + 0.005Ω)± (0.035% reading + 0.005Ω)1mΩ
6 Types of Thermal Resistance °CSee “RTD Thermal Resistance Technical Indicators”0.1°C
50V Measurement Input Resistance: ≥ 1MΩ; 5V Measurement Input Resistance: ≥ 500MΩ; Current Measurement Input Resistance: ≤ 10Ω; Resistance, Thermal Resistance Measurement Current: 500Ω/1mA; 5000Ω/0.5mA; Preheating Time: 30 minutes; Working Temperature: -10 ~ +50°C; Relative Humidity: 0 ~ 90% RH; Storage Temperature: -20 ~ +60°C; Temperature Coefficient: ± (0.0002 reading + 0.0002% range) / °C (< 15°C or > 25°C)

 

 

  • TC Thermocouple Technical Indicators (One Year, Ambient Temperature: 20 ± 5°C)

 

Thermocouple GraduationTemperature Range (°C)Max. Error Allowed (Simulation Output and Measurement)
0.01 Grade0.02 Grade0.05 Grade
S(-20 ~ 200)± 0.9°C± 0.9°C± 0.9°C
(200 ~ 700)± 0.5°C± 0.5°C± 0.6°C
(700 ~ 1500)± 0.5°C± 0.5°C± 0.8°C
(1500 ~ 1768)± 0.5°C± 0.6°C± 1.0°C
R(-20 ~ 100)± 0.9°C± 0.9°C± 0.9°C
(100 ~ 600)± 0.6°C± 0.6°C± 0.6°C
(600 ~ 1600)± 0.4°C± 0.5°C± 0.7°C
(1600 ~ 1768)± 0.5°C± 0.6°C± 1.0°C
J(-200 ~ 100)± 0.3°C± 0.3°C± 0.4°C
(-100 ~ 300)± 0.2°C± 0.2°C± 0.3°C
(300 ~ 900)± 0.2°C± 0.2°C± 0.4°C
(900 ~ 1200)± 0.2°C± 0.3°C± 0.5°C
K(-250 ~ -150)± 1.0°C± 1.1°C± 1.3°C
(-150 ~ 1000)± 0.2°C± 0.2°C± 0.4°C
(1000 ~ 1372)± 0.3°C± 0.4°C± 0.7°C
T(-250 ~ -200)± 0.8°C± 0.9°C± 1.0°C
(-200 ~ 0)± 0.3°C± 0.4°C± 0.4°C
(0 ~ 400)± 0.2°C± 0.2°C± 0.2°C

 

Thermocouple GraduationTemperature Range (°C)Max. Error Allowed (Simulation Output and Measurement)0.01 Grade0.02 Grade0.05 Grade
N(-200 ~ -100)±0.5°C±0.5°C±0.6°C
(-100 ~ 600)±0.3°C±0.3°C±0.3°C
(600 ~ 1300)±0.3°C±0.4°C±0.6°C
E(-250 ~ -200)±0.5°C±0.5°C±0.8°C
(-200 ~ 700)±0.3°C±0.3°C±0.5°C
(700 ~ 1000)±0.2°C±0.3°C±0.6°C
B(600 ~ 800)±0.8°C±0.8°C±1.2°C
(800 ~ 1100)±0.5°C±0.8°C±1.1°C
(1100 ~ 1820)±0.5°C±0.6°C±0.9°C
WRe3-25(0 ~ 1800)±0.5°C±0.8°C±1.1°C
(1800 ~ 2200)±0.7°C±1.0°C±1.5°C
(2200 ~ 2315)±0.8°C±1.2°C±1.7°C
WRe5-26(0 ~ 1000)±0.5°C±0.7°C±1.0°C
(1000 ~ 1800)±0.7°C±0.9°C±1.3°C
(1800 ~ 2315)±0.8°C±1.1°C±1.9°C

Notes

  • Meet 90 international temperature standard, determined by max. error allowed for mV output and measurement.
  • Exclude sensor and compensating wire error.
  • With precision for external cold joint, add 0.2°C for internal joint.
  • Max. error allowed for Pt100 temperature measurement probe: ±0.2°C.